5 calibration, Calibration – YSI pHotoFlex pH Handheld Colorimeter User Manual
Page 53
pHotoFlex
®
pH
Operation
53
ba76123e01
02/2013
4.6.5
Calibration
Why calibrate?
pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is dis-
played. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the meter. Thus, you
should calibrate at regular intervals.
When to calibrate?
After connecting another electrode
When the sensor symbol flashes:
– after the calibration interval has expired
– after voltage interruption (e.g. empty batteries, empty
rechargeable battery)
Buffer sets for
calibration
You can use the buffer sets quoted in the table for an automatic calibra-
tion. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into account during
calibration.
The buffers are selected in the Configuration / pH & ORP / Cal. type
menu, see section 4.6.4).
Buffer set
Name on the
display
pH values at
25 °C
YSI technical buffer solutions
TEC
2.00
4.01
7.00
10.01
NIST/DIN buffer solutions
NIST/DIN
1.679
4.006
6.865
9.180
12.454