beautypg.com

5 calibration, Calibration – YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 53

background image

pHotoFlex

®

pH

Operation

53

ba76123e01

02/2013

4.6.5

Calibration

Why calibrate?

pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is dis-
played. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the meter. Thus, you
should calibrate at regular intervals.

When to calibrate?

 After connecting another electrode
 When the sensor symbol flashes:

– after the calibration interval has expired

– after voltage interruption (e.g. empty batteries, empty

rechargeable battery)

Buffer sets for

calibration

You can use the buffer sets quoted in the table for an automatic calibra-
tion. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into account during
calibration.

The buffers are selected in the Configuration / pH & ORP / Cal. type
menu, see section 4.6.4).

Buffer set

Name on the
display

pH values at
25 °C

YSI technical buffer solutions

TEC

2.00

4.01

7.00

10.01

NIST/DIN buffer solutions

NIST/DIN

1.679

4.006

6.865

9.180

12.454