YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 57

background image

pHotoFlex

®

pH

Operation

57

ba76123e01

02/2013

For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the asymmetry of the elec-
trode.

5

Using <> <>, select Continue and press TER>. The buffer is measured.

The measured value is checked for stability (AutoRead).

6

Wait for the end of the AutoRead measurement or accept the
calibration value with .
The calibration display for the next buffer appears.

7

For single-point calibration, select Exit with one point calibra-
tion
with <> <> and confirm with .

The calibration is completed as a single-point calibration.

The new calibration values are displayed as an informative
message (

i

).

You have the following options:
 Accept the new calibration values with .

Subsequently, the calibration record is displayed and output
to the interface at the same time.

 To switch to the measured value display without accepting

the new calibration values, press (short pressure) or
.

i Buffer value = 7.00
i U = 3 mV
i Temperature =

24.8 °C

Terminate AutoRead

pH & ORP \ Calibration

i Buffer recognition TEC
i Immerse sensor in buffer 2

Exit with one point calibration

Continue

pH & ORP \ Calibration