YSI pHotoFlex pH Handheld Colorimeter User Manual
Page 59
pHotoFlex
®
pH
Operation
59
ba76123e01
02/2013
Continuing for three-
point calibration
(Cal. type TEC)
14
Thoroughly rinse the electrode with distilled water.
15
Immerse the electrode in buffer solution 3.
16
If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with <> <> and
17
Using <> <>, select Continue and press
The measured value is checked for stability (AutoRead).
18
Wait for the end of the AutoRead measurement or Terminate
AutoRead with
value.
The new calibration values are displayed as an informative
message (
i
).
You have the following options:
Accept the new calibration values with
Subsequently, the calibration record is displayed and output
to the interface at the same time.
To switch to the measured value display without accepting
the new calibration values, press
i Buffer value = 4.01
i U = 184 mV
i Temperature =
24.8 °C
Terminate AutoRead
pH & ORP \ Calibration
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (121 pages)
- 5200A (128 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (143 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IFL 700 IQ (46 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)