YSI pHotoFlex pH Handheld Colorimeter User Manual
Page 55
pHotoFlex
®
pH
Operation
55
ba76123e01
02/2013
Sample of a record
Calibration evaluation
After calibrating, the meter automatically evaluates the calibration. The
asymmetry and slope are evaluated separately. The worse evaluation
of both is taken into account. The evaluation appears on the display
and in the calibration record.
31.10.03 16:13
pHotoFlex Ser. no. 12345678
Calibration pH & ORP
Calibration date
31.10.03 16:13:33
Calibration interval
7 d
AutoCal TEC
Buffer 1 4.01
Buffer 2 7.00
Buffer 3 10.01
Voltage 1 184.0 mV 24.0 °C
Voltage 2 3.0 mV 24.0 °C
Voltage 3 -177.0 mV 24.0 °C
Slope
-60.2 mV/pH
Asymmetry 4.0 mV
Sensor +++
Sample dis-
play
Calibration
record
Asymmetry
[mV]
Slope
[mV/pH]
+++
-15 ... +15
-60.5 ... -58
++
-20 ... +20
-58 ... -57
+
-25 ... +25
-61 ... -60.5
or
-57 ... -56
-
-30 ... +30
-62 ... -61
or
-56 ... -50
Clean the electrode according
to the electrode operating
manual
----
----
< -30 or
> 30
... -62 or
... -50
Eliminate the error according to
chapter 6 W
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (128 pages)
- 5200A (121 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9500 Photometer (143 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IFL 700 IQ (46 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)