Operation

pHotoFlex

®

pH

66

ba76123e01

02/2013

Output format, CSV

0;01.12.11;14:18:07;1;81;O2;[BV];[2 + 1];166;mg/l;VALID;;;
14;03.12.11;07:41:51;3;;;;;24.8;%;VALID;557;nm;VALID

2

1

3

4 5

6

7

8

9

12

10

13

11

14

Data

Description

1

No.

Current number of the storage location (or "0" during output from mea-
sured value display)

2

Date

Date of storing

3

Time

Time of storing

4

ID

adjusted ID

5

Program number

Only for the measured parameter, Concentration

6

Citation form

Only for the measured parameter, Concentration

7

AR/

BV

 Measuring mode, pH & ORP: AutoRead
 Measured parameter, Concentration: Blank value

8

Dilution

Only for the measured parameter, Concentration

9

Measured value

 Measured value or
 Upper/lower measuring range limit

(only with measured value status, OFL/UFL)

10

Unit of 9

 Unit of the measured value or
 Designation of dimensionless measured values, e.g.

11

Measured value
status of 9

 VALID: Measured value valid
 INVALID: Measured value invalid
 UFL: Measured value below the lower measuring range limit
 OFL: Measured value above the upper measuring range limit

12

Secondary mea-
sured value

or

Standard adjustment

 Measuring mode, pH & ORP: Temperature
 Measured parameter, Absorbance / % Transmission: Wavelength
 for the measured parameter, Concentration: Status standard adjust-

ment [Cal]

13

Unit of 12

 Measuring mode, pH & ORP: °C / °F
 Measured parameter, Absorbance / % Transmission: nm

14

Measured value sta-
tus of 12

VALID, INVALID, UFL, OFL (see point 11)