beautypg.com

2 info messages, Info messages -2 – YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual

Page 50

background image

Indexes

TriOxmatic

®

702 IQ

9 - 2

ba76048e01

01/2012

9.1.2

Info messages

EI1333

Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components

EI2333

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components

ES1333

Component hardware defective
* Contact service

ESB333

SensReg: Electrolyte supply is depleted
* Change electrolyte solution and membrane head immediately (see
operating manual)

Message code

Message text

Message code

Message text

IC1333

Sensor has been successfully calibrated
* For calibration data, see calibration history

IC4333

Last valid calibration has been activated. Make sure the sensor
operates correctly.

II1333

Language not available,
Default language German
* Contact service