9.1.2
Info messages
EI1333
Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components
EI2333
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components
ES1333
Component hardware defective
* Contact service
ESB333
SensReg: Electrolyte supply is depleted
* Change electrolyte solution and membrane head immediately (see
operating manual)
Message code
Message text
Message code
Message text
IC1333
Sensor has been successfully calibrated
* For calibration data, see calibration history
IC4333
Last valid calibration has been activated. Make sure the sensor
operates correctly.
II1333
Language not available,
Default language German
* Contact service