beautypg.com

9 indexes, 1 explanation of the messages, 1 error messages – YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual

Page 49: Indexes -1, Explanation of the messages -1 9.1.1, Error messages -1, 9indexes

background image

TriOxmatic

®

702 IQ

Indexes

9 - 1

ba76048e01

01/2012

9

Indexes

9.1

Explanation of the messages

This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S

ENSOR

N

ET

system for the TriOxmatic

®

702 IQ sensor.

Note
Information on the contents and structure of the log book, and how to
call it up, is given in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system operating manual.

Note
All message codes of the TriOxmatic

®

702 IQ end up with "333".

9.1.1

Error messages

Message code

Message text

EA1333

Meas. range exceeded or undercut
* Check process
* Select other meas. range

EA2333

Sensor temperature too high!
* Check process and application

EA3333

Sensor temperature too low!
* Check process and application

EC4333

Sensor could not be calibrated,
sensor blocked for measurement
Cause: instable signal
* Check polarization time and temperature adjustment
* Check calibration conditions
* Read calibration history
* Repeat calibration, if unsuccessful: service the sensor (see operating
manual)

EC5333

Sensor could not be calibrated,
sensor blocked for measurement
Cause: slope error
* Check polarization time and temperature adjustment
* Check calibration conditions
* Read calibration history
* Repeat calibration, if unsuccessful: service the sensor (see operating
manual)