YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual
Page 15
TriOxmatic
®
702 IQ
Commissioning
3 - 5
ba76048e01
01/2012
Carrying out settings
Using
s
, switch from the measured value display to the main menu of
the settings. Then navigate to the setting menu (setting table) of the
sensor. The exact procedure is given in the relevant IQ S
ENSOR
N
ET
system operating manual.
Temp. adjustment
-1.5 K ... +1.5 K
The temperature compensation enables the
tolerances of the temperature sensor to be
balanced (shifting of the zero point by
±1.5 K).
Notes:
Due to the thermal capacity of the sensor,
it is necessary to place it in a container
with at least 2 liters of water.
Leave the sensor in this container for at
least 15 minutes while stirring
occasionally until the balancing can be
performed.
With temperature differences between
the water and sensor > 10 °C, leave the
sensor for at least 1 hour in this container
while stirring occasionally until the
balancing can be performed.
Save and quit
The system confirms the saving of the
settings and the display switches to the next
higher level.
Quit
The display switches to the next higher level
without saving the new settings.