4 trioxmatic® 702 iq setting table, Trioxmatic, Sensor (see section 3.4) – YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual
Page 14
Commissioning
TriOxmatic
®
702 IQ
3 - 4
ba76048e01
01/2012
3.4
TriOxmatic
®
702 IQ setting table
Setting
Selection/values
Explanation
Measuring mode
Concentration
Saturation
Unit of the measured value on the measured
value display.
Measuring range
Concentration
0 ... 2000 µg/l
0.0 ... 15 mg/l
0 ... 2000 ppb
0 ... 10.00 ppm
These measuring ranges are available for
selection.
Measuring range
Saturation
0 ... 110.0 %
The measuring range is set permanently.
Calibration
(software version
2.21 or higher)
valid
invalid
last valid
active
abort
Displays and determines which calibration
data the measured value calculation is
based on. The active calibration is displayed
in the calibration history (see section 4.2.3).
valid indicates that a valid calibration is
available. The value cannot be changed.
invalid is displayed if the last calibration is
invalid and the sensor is blocked for measu-
rement. In this case, you can switch to the
values of the last valid calibration, provided a
valid calibration is available in the sensor.
Thus you activate the last valid calibration
stored in the sensor the next time you exit
the setting table with Save and quit. The next
time the setting table is opened, valid is dis-
played.
active indicates that the sensor is being cali-
brated. If Cancel is selected, the active cali-
bration procedure is canceled as soon as the
setting table is exited with Save and quit.
Temperature mode
°C
°F
Unit of the measured temperature value
(Celsius, Fahrenheit).