Maintenance, cleaning, disposal, and replacement

TriOxmatic

®

702 IQ

5 - 14

ba76048e01

01/2012

12

Carefully shake off the drops of water.

13

Fill a new WP 25i membrane cap and screw it on (see section
5.3 C

HANGING

THE

ELECTROLYTE

AND

MEMBRANE

CAP

).

14

Connect the sensor again with the SACIQ sensor connection
cable.

15

After approx. 60 minutes, the sensor is ready for operation.

16

Recalibrate the sensor (see section 4.2 C

ALIBRATION

).