5 check the sensor for freedom from zero current, Check the sensor for freedom from zero current -15 – YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual

Page 37

background image

TriOxmatic

®

702 IQ

Maintenance, cleaning, disposal, and replacement

5 - 15

ba76048e01

01/2012

5.5

Check the sensor for freedom from zero current

The sensor is zero-current free. Therefore, no calibration of the sensor
in an oxygen-free solution is included.

However, it is possible to check the freedom from zero-current
according to DIN EN 25814/ISO 5814. This is advisable if there is any
suspicion of a malfunction.

Test solution

1 g/l aqueous sodium sulfite solution, Na

2

SO

3

(addition of 1 mg/l of a cobalt(II) salt accelerates the removal of oxygen
from the solution.)

The sensor should be in operation for at least 1 hour before the
inspection.

Test procedure

Evaluation

The sensor is OK if the terminal shows < 1 % oxygen saturation.

If the measured value is > 1 %:

 Clean the gold working electrode (see section 5.4.1 C

LEANING

THE

GOLD

WORKING

ELECTRODE

) and change the electrolyte and

membrane cap (see section 5.3 C

HANGING

THE

ELECTROLYTE

AND

MEMBRANE

CAP

).

 If necessary, also clean the silver counter electrode (see section

5.4.2 C

LEANING

THE

SILVER

COUNTER

ELECTRODE

).

1

Clean the sensor (see section 5.2 C

LEANING

THE

SENSOR

SHAFT

AND

MEMBRANE

).

2

In the sensor setting table, select the measuring mode %
(unit %) and switch to the measured value display with

m

.

3

Submerse the sensor in the test solution.

4

Leave the sensor in the test solution for an hour.

5

Read the measured value.