Maintenance, cleaning, disposal, and replacement

TriOxmatic

®

702 IQ

5 - 4

ba76048e01

01/2012

Changing the electrolyte

and membrane cap

Caution

The ELY/A electrolyte solution irritates eyes, skin and mucus
membranes. If it comes into contact with the eyes, rinse
thoroughly with water and consult a doctor! During working
activities, always wear suitable protective gloves and protective
goggles/face shield! Follow the safety datasheet.

1

Take the sensor out of the flow-thru vessel.

2

Unscrew the membrane head (for disposal of the membrane
head and electrolyte solution, see section 5.7).

3

Rinse the sensor head with tap water.