Technical data
TriOxmatic
®
702 IQ
7 - 4
ba76048e01
01/2012
7.5
Characteristic data on delivery
Zero signal
< 1 ppb (0.001 mg/l)
Response time at 25 °C
t
90
(90 % of the final value display after) < 30 s
t
99
(99 % of the final value display after) < 110 s
Adjustment time
Adjustment time for measuring low oxygen concentrations after air
calibration:
Adjustment time for measuring low oxygen concentrations after
exchanging the membrane head and electrolyte solution and
subsequent air calibration:
Own consumption
0.34 µg h
-1
(mg/l)
-1
at 20 °C
Drift
< 1 % per month during permanent polarization
Working life per
electrolyte filling
D. O. concentration
Adjustment time
10 ppb
10 min
5 ppb
15 min
< 2 ppb
45 min
D. O. concentration
Adjustment time
10 ppb
40 min
5 ppb
70 min
< 2 ppb
180 min
D. O. concentration
Theoretical electrolyte reserve
Air saturation
approx. 45 days
< 100 ppb
approx. 12 years
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (121 pages)
- 5200A (128 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (143 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IFL 700 IQ (46 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)