beautypg.com

5 characteristic data on delivery, Characteristic data on delivery -4 – YSI IQ S ENSOR N ET TriOxmatic 702 IQ User Manual

Page 46

background image

Technical data

TriOxmatic

®

702 IQ

7 - 4

ba76048e01

01/2012

7.5

Characteristic data on delivery

Zero signal

< 1 ppb (0.001 mg/l)

Response time at 25 °C

t

90

(90 % of the final value display after) < 30 s

t

99

(99 % of the final value display after) < 110 s

Adjustment time

Adjustment time for measuring low oxygen concentrations after air
calibration:

Adjustment time for measuring low oxygen concentrations after
exchanging the membrane head and electrolyte solution and
subsequent air calibration:

Own consumption

0.34 µg h

-1

(mg/l)

-1

at 20 °C

Drift

< 1 % per month during permanent polarization

Working life per

electrolyte filling

D. O. concentration

Adjustment time

10 ppb

10 min

5 ppb

15 min

< 2 ppb

45 min

D. O. concentration

Adjustment time

10 ppb

40 min

5 ppb

70 min

< 2 ppb

180 min

D. O. concentration

Theoretical electrolyte reserve

Air saturation

approx. 45 days

< 100 ppb

approx. 12 years