background image

HandyLab 100

What to do if...

55

ba77062d01

03/2014

6

What to do if...

Error message

OFL, UFL

Error message,

Error

Cause

Remedy

pH electrode:

– Measured value outside the

measuring range

– Use suitable electrode

– Air bubble in front of the junc-

tion

– Remove air bubble

– Air in the junction

– Extract air or moisten junc-

tion

– Cable broken

– Replace the electrode

– Gel electrolyte dried out

– Replace the electrode

Cause

Remedy

pH electrode:

– The values determined for

zero point and slope of the
electrode are outside the
allowed limits.

– Recalibrate

– Junction contaminated

– Clean the junction

– Electrode broken

– Replace the electrode

Buffer solutions:

– Incorrect buffer solutions

– Change calibration proce-

dure

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Change solutions