Continuing with two-
point calibration
Continuing with three-
to five-point calibration
11
Thoroughly rinse the electrode with deionized water.
12
Immerse the electrode in buffer solution 2.
13
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <
><>.
14
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
15
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with
The calibration display for the next buffer appears.
16
Set the nominal buffer value for the measured temperature with
<
><>.
17
Accept the calibration value with
The calibration display for the next buffer appears (voltage dis-
play).
18
Finish the calibration procedure as a two-point calibration with
The calibration record is displayed.
pH
4.035
24.8 °C
VariCal
2
15.03.2014
010:00
Buffer
19
Thoroughly rinse the electrode with deionized water.
20
Immerse the electrode in the next buffer solution.
21
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <
><>.