beautypg.com

8 carrying out a manual calibration (varical), Carrying out a manual calibration (varical) – Xylem HandyLab 100 User Manual

Page 41

background image

HandyLab 100

Operation

41

ba77062d01

03/2014

4.4.8

Carrying out a manual calibration (VariCal)

Make sure that in the sensor menu, Buffer menu, the

VariCal buffer set

is correctly selected (see page 30).

Use any one to five buffer solutions in ascending or descending order.

Note
If single-point calibration was set in the menu, the calibration procedure
is automatically finished with the measurement of buffer solution 1 and
the calibration record is displayed.

1

In the measured value display, select the measured parameter
pH or mV with .

2

Start the calibration with .
The calibration display appears.

3

Thoroughly rinse the electrode with deionized water.

4

Immerse the electrode in buffer solution 1.

5

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

><>.

6

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.

pH

0.0

24.8 °C

VariCal

1

15.03.2014

010:00

Buffer

mV