Xylem HandyLab 100 User Manual

Page 27

background image

HandyLab 100

Operation

27

ba77062d01

03/2014

Criteria for a stable

measured value

The Stability control function checks whether the measured values are
stable within the monitored time interval.

The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.

Note
You can prematurely terminate the Stability control function manually
with at any time. If the Stability control function is prema-
turely terminated, the current measurement data are output to the
interface without the AutoRead info.

3

Using , start a further measurement with stability
control.
or
Release the frozen measured value again with or .
The [AR] status display disappears. The display switches back
to the previous indication.

Measured
parameter

Time interval

Stability during the time
interval

pH value

15 seconds

Δ

: better than 0.01 pH

Temperature

15 seconds

: better than 0.5 °C