2 measuring the ph value, Measuring the ph value – Xylem HandyLab 100 User Manual
Page 26
Operation
HandyLab 100
26
ba77062d01
03/2014
4.4.2
Measuring the pH value
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of
the setting for automatic Stability control (see page 24) in the Measure-
ment menu.
1
Perform the preparatory activities according to section 4.4.1.
2
Immerse the pH electrode in the test sample.
3
Select the pH or mV display with
pH
6.949
25.0
°C
15.03.2014
010:00
Menu
1
Freeze the measured value with
The [HOLD] status indicator is displayed.
The HOLD function is active.
2
Using
ally.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized.