Operation
HandyLab 100
30
ba77062d01
03/2014
4.4.4
Settings for pH and ORP measurements
Overview
The following settings are possible for pH and ORP measurements:
Resolution
Calibration interval
Buffers for calibration
Unit of the temperature
Automatic stability control
Unit for slope
Calibration record (display)
Settings
The settings are made in the menu for calibration and measurement
settings of the pH/ORP measurement. To open the settings, display the
required parameter in the measured value display and press the
[menu] or
measured value display with
Menu item
Possible
setting
Explanation
Calibration / Calibra-
tion record
-
Displays the calibration
record of the last calibration.
Calibration / Buffer
TEC
NIST/DIN
VariCal
...
Buffer sets to be used for pH
calibration.
More buffers and details,
see section 4.4.5.
Calibration / One
point calibration
Yes
No
Quick calibration with 1 buf-
fer
Calibration / Calibra-
tion interval
1 ... 999 d
Calibration interval for the
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring screen.
Calibration / Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the
Nernst slope of
-59.2 mV/pH (100 x deter-
mined slope/Nernst slope).
Man. temperature
-25 ... +130 °C
Entry of the manually deter-
mined temperature. For
measurements without tem-
perature sensor only.