Xylem HandyLab 100 User Manual
Page 38
Operation
HandyLab 100
38
ba77062d01
03/2014
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the electrode.
Continuing with two-
point calibration
6
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
7
Wait for the end of the measurement with stability control or
accept the calibration value with
The calibration display for the next buffer appears (voltage dis-
play).
8
If necessary, finish the calibration procedure as a single-point
calibration with
The calibration record is displayed.
pH
-180.0
24.8 °C
AutoCal TEC
10.011
15.03.2014
010:00
AR
Buffer
mV
9
Thoroughly rinse the electrode with deionized water.
10
Immerse the electrode in buffer solution 2.
11
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <
><>.
12
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.