Setting up the cell

constant manually

To set up the cell constant manually, proceed as follows:

Note

1 Press the key repeatedly until LF CELL ap-

pears on the display.

2 Press the key.

3 Press the key repeatedly until the cell con-

stant to be set, e. g. 0.475 cm

-1

is displayed.

4 Set up the cell constant to be used by pressing

<

> <

>, e. g. 0.614 cm

-1

.

5 To return to the measuring mode: press the <

ì