Xylem CG 853 User Manual
Page 31
Operation
105
Note
This method of automatically determining the cell constant
by calibrating in the 0.01 mol/l KCL control standard can
only be used for measuring cells with a cell constant in the
range 0.450 ... 0.500 cm
-1
or 0.800 ... 1.200 cm
-1
.
Calibration
evaluation
After calibrating, the instrument automatically evaluates the
current condition of the calibration.
The evaluation appears on the display.
Calibration protocol
The calibration protocol contains the calibration data of the
current calibration.
Note
You can automatically print out a calibration protocol after
the calibration. To do so, connect a printer to the interface
according to section 4.5.3 before calibrating. After a valid
calibration, the protocol is printed.
Display
Cell constant [cm
-1
]
0.450 ... 0.500 cm
-1
0.800 ... 1.200 cm
-1
Perform error elimination
according to chapter
6 W
HAT
TO
DO
IF
...
outside of the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 1.200 cm
-1
CALIBRATION PROTOCOL
14.04.99 11:37
Device No.: 99990000
CALIBRATION CONDUCTIVITY
Cal Time: 14.04.99 / 11:37
Cal Interval: 180d
Cal Std.: 0.01 mol/l KCL
40.0 °C
Conduct./Tref25: 1413µS/cm
Cell Const : 0.975 1/cm
Probe : +++