84

Given as measuring uncertainty at a confidence level of
95%. All accuracies additionally depend on the accuracies
of the measuring probes.

No

compensation:

Accuracy

± 0.5 %

Non-linear compensation

:

Accuracy

Test sample temperature

± 0.5 %

0 °C ... 35 °C

according to EN 27 888;
± 0.5 %

35 °C ... 50 °C

extended nLF function according to
SCHOTT measurements

Linear compensation

:

Accuracy

Test sample temperature

± 0.5 %

10 °C ... 75 °C

(the accuracy percentage always refers
to the measured value!)

SAL

Range 0.0 ... 42.0
Accuracy

Test sample temperature

± 0.1

at 5 °C ... 25 °C

± 0.2

at 25 °C ... 30 °C

TDS [mg/l]

1

T [°C]

NTC 30:
Accuracy± 0.1

PT 1000:
Accuracy

Operating temperature

± 0.5

0 °C ... 15 °C

± 0.1

15 °C ... 35 °C

± 1

35 °C ... 55 °C

Measuring accuracy