Xylem CG 853 User Manual

Page 30

background image

Operation

104

Note
If error message

appears see chapter 6 W

HAT

TO

DO

IF

...

AutoRead

During calibrating, the

AutoRead

function is automatically

activated. The AR display flashes. The calibration procedure
is finished when the AR display stops flashing.

4 Immerse the measuring cell into the 0.01 mol/l KCL

control standard.

5 Press the

key.

– If no temperature probe is connected, enter the

current temperature of the solution using <

> <

>

and confirm with

– If a temperature probe is connected, the AR

measurement to determine the cell constant starts.

flashes until a stable signal is achieved. The

determined cell constant ist displayed; the
measuring instrument automatically stores the cell
constant.