beautypg.com

Xylem CG 853 User Manual

Page 24

background image

Operation

98

AutoRead AR

(Drift control)

The

AutoRead

function (drift control) checks the stability of

the measurement signal. The stability has a considerable ef-
fect on the reproducibility of the measured value.

Note
The current AutoRead measurement (with acceptance of
the current value) can be terminated at any time by pressing
.

1 Call up the measuring mode by pressing <

ì

>.

2 Activate the AutoRead function by pressing

read>.
The current measured value is frozen (Hold function).

3 Start the AutoRead function by pressing

.
AR flashes on the display until a stable measured val-
ue is reached.
This measured value is transmitted to the interface.

4 If necessary, start the next AutoRead measurement

by pressing .

5 To terminate the AutoRead function: Press the

key.