Operation

100

4.2.2 Salinity

To measure the salinity, proceed as follows:

1 Perform the preparatory activities according to sec-

tion 4.2.

2 Immerse the conductivity measuring cell into the test

sample.

3 Press the <

ì

> key repeatedly until Sal appears in the

status line. The salinity value of the sample appears
on the display.

4 If necessary, set the temperature by pressing

<

> <

> (see section 4.2, page 97).

5 Wait for a stable measured value.

6 Perform an AutoRead measurement according to

section 4.2.