7 Spare Parts


Recommended spare parts for the GPR-2800AIS Oxygen Transmitter:

Item No.

Description

GPR-11-24

Oxygen Sensor

XLT-11-24

Oxygen Sensor


Other spare parts:

Item No.

Description

HTR-1004

Heater Low Power Sensor Housing

A-1004-2-14

Housing Sensor Stainless Steel

A-1016-A

Housing Sensor Bottom Assembly Stainless Steel

B-2762-A-2-14

Housing Sensor Upper Assembly Stainless Steel

MTR-1011

Meter Digital Panel LCD Backlight

MTR-1014

Meter Digital Panel LCD Low Temperature

ORNG-1007

O-ring 3/32 x 1-3/8 x 1-9/16 Viton

A-1151-AIS-2

PCB Assembly Main / Display

A-1166-AIS-AC

PCB Assembly Interconnection / AC Power Supply

A-1166-AIS-DC

PCB Assembly Interconnection / DC Power Supply


8 Troubleshooting


Symptom

Possible Cause

Recommended Action


Slow recovery or










At installation, defective sensor

Air leak in sample system connection(s)



Abnormality in zero gas
Damaged in service - prolonged exposure to
air, electrolyte leak
Sensor nearing end of life


Replace sensor if recovery unacceptable or O

2

reading fails to reach 10% of lowest range
Leak test the entire sample system: Vary the flow
rate, if the O

2

reading changes inversely with the

change in flow rate indicates an air leak - correct
source of leak
Qualify zero gas (using portable transmitter)
Replace sensor

Replace sensor


High O

2

reading

after installing
or replacing sensor


Transmitter calibrated before sensor stabilized
caused by:
1) Prolonged exposure to ambient air, worse
if sensor was unshorted
2) Air leak in sample system connection(s)
3) Abnormality in zero gas


Allow O

2

reading to stabilize before making the

span/calibration adjustment
Continue purge with zero gas

Leak test the entire sample system (above)
Qualify zero gas (using portable transmitter)


High O

2

reading

Sampling


Flow rate exceeds limits
Pressurized sensor
Improper sensor selection


Correct pressure and flow rate
Remove restriction on vent line
Replace GPR/PSR sensor with XLT sensor when
CO

2

or acid gases are present