Advanced instruments inc – Analytical Industries GPR-2800 AIS ATEX Oxygen Analyzer User Manual

Page 24

background image

Advanced Instruments Inc.



24



FACTORY

DEFAULTS

SET




3.3

%

AUTO SAMPLING

10% RANGE

24.5 C 100 KPA
LO1 2% 10% HI2



Output Zero
In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. This feature enables the user to
adjust the 4mA signal output when the LCD displays 00.00. Note: Adjust the 20mA signal output with the OUTPUT SPAN option
described below.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.

4. The following displays appear:

MAIN MENU

AUTO SAMPLE
MANUAL SAMPLE

CALIBRATION

CONFIG ALARMS
BYPASS ALARMS



>>>

CALIBRATION

SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
OUTPUT SPAN

OUTPUT ZERO



5. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
6. Press the ENTER key to select the highlighted menu option and the following display appears:

100.0

OUTPUT ZERO OFFSET

PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN





7. Compute the adjustment value as described in Appendix B or consult the factory. The true adjustment value must be

determined empirically by trial and error. Adjust the initial adjustment value for additional percent errors.


090.0

OUTPUT ZERO OFFSET

PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN