2 quality control certification, Advanced instruments inc – Analytical Industries GPR-2800 AIS ATEX Oxygen Analyzer User Manual
Page 3

Advanced Instruments Inc.
3
2 Quality Control Certification
Date:
Customer: Order No.:
Pass
Model:
GPR-2800AIS Oxygen Transmitter S/N __________________
Sensor:
( ) GPR-11-24 Oxygen Sensor
( ) XLT-11-24 Oxygen Sensor S/N __________________
Accessories:
Owner’s Manual
Configuration:
A-1151-AIS-2 PCB Assembly Main Processing Software Ver: ___________
Power: ( ) A-1166-AIS-AC PCB Assembly Alarms/AC Power 110V & Interconnection
( ) A-1166-AIS-DC PCB Assembly Alarms/DC Power 9-28V & Interconnection
Ranges: 0-1%, 0-5%, 0-10%, 0-25%
A-1004-2-14 Stainless steel sensor housing, 1/8” tube fittings
Barometric pressure and temperature compensation
NEMA 4X rated wall mount enclosures
Backplate 10 x 16” suitable for wall mounting
Test: Pass
Pass
Set default zero
Noise level <
± 1.0% FS
Set default span @ 50uA
Alarm delay
Zero calibration
Alarm bypass
Span Calibration
Alarm configurations; ALARM 1, ALARM 2
Analog signal output 4-20mA full scale
Alarm function; ALARM 1, ALARM 2
Calibrates with adequate span adjustment within 10-50% FS
Alarm relays; ALARM 1, ALARM 2
Baseline drift on zero gas <
± 2% FS over 24 hour period
Alarm system fail, dry contact
Overall inspection for physical defects
Options:
Notes: