Advanced instruments inc – Analytical Industries GPR-2800 AIS ATEX Oxygen Analyzer User Manual

Page 20

background image

Advanced Instruments Inc.



20

MAIN MENU

AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION

CONFIG ALARMS

BYPASS ALARMS



>>>

MAIN MENU

SET ALARM 1
SET ALARM 2
SET ALARM DELAY

ALARM 1 HI/LO

ALARM 2 HI/LO

ALARMS AUDIBLE/SILENT



5. Advance the reverse shade cursor using the ARROW keys to highlight the ALARM 1 option, which appears as either ALARM

1 HI or ALARM 1 LO.

6. Press the ENTER key to toggle and change the displayed setting. After 3 seconds, the system returns to SAMPLING mode.


3.3

%

AUTO SAMPLING

10% RANGE

24.5 C 100 KPA
LO1 2% 10% HI2



7. Repeat steps 1 through 6 for the ALARM 2 HI/LO setting.

Set Local Alarms:

1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CONFIG ALARMS.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appears:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE
CALIBRATION

CONFIG ALARMS

BYPASS ALARMS



>>>

MAIN MENU

SET ALARM 1

SET ALARM 2
SET ALARM DELAY
ALARM 1 HI/LO
ALARM 2 HI/LO

ALARMS AUDIBLE/SILENT



5. Advance the reverse shade cursor using the ARROW keys to highlight the ALARMS AUDIBLE/SILENT option, which appear

as either ALARMS AUDIBLE or ALARMS SILENT.

6. Press the ENTER key to toggle and change the displayed setting. After 3 seconds, the system returns to SAMPLING mode.