beautypg.com

Wiring and configuration of the ecg – OSRAM OPTOTRONIC Constant current LED power supplies with 3DIM User Manual

Page 13

background image

13

WIRING AND CONFIGURATION OF THE ECG

Figure 18a shows the setup if only 3DIM ECGs are con-
nected to the SD control switch. In case other ballasts
or loads are supplied via the SD line (e.g. L2) or an exter-
nal surge protection device is connected to this port, an
additional relay needs to be used within each luminaire
(see fi gure 18b), unless something different is mentioned
in the product- related documents (data sheet or instruc-
tion sheet). The used relay should be suitable for low
switching currents < 1 mA (mostly gold plated contacts).

The SD control switch shown in fi gures 18a/b can be a re-
lay connected to a timer with a remote HIGH/LOW voltage
signal.

For details, please refer to the 3DIM Tool manual.

Figure 18a: StepDIM activation for a group of 3DIM ECGs by the same control switch (example).

Figure 18b: StepDIM port set-up inside luminaire for mixed installation (Phase L2 is used to activate SD and also to supply other ballasts or loads).

Luminaire 1

Luminaire 1

Depending on
the ECG type

Depending on
the ECG type

3DIM ECG

3DIM ECG

Relay

Relay

LED out

LED out

L

L

L

L

L

N

N

N

N

N

SD

SD

SD

SD

SD

DA

DA

DA

DA

DA

DA

DA

DA

DA

DA

LED out

LED out

LED out

3DIM ECG

3DIM ECG

3DIM ECG

ECG 1

ECG 2

Depending on
the ECG type

Depending on
the ECG type

Depending on
the ECG type

Light
load

Light
load

Light
load

Light
load

Light
load

Light
load

Light
load

Luminaire 2

Luminaire 2

Luminaire 3

Luminaire 3

SD
control switch

SD
control switch

L1

L1

N

N

L2

L2

PE

PE