Nirquest gratings – Ocean Optics Nirquest Install User Manual
Page 38
![background image](/manuals/292632/38/background.png)
B: Specifications
32
016-700000-000-02-201301
NIRQuest Spectrometer – Slit Size vs. Resolution
and Throughput
The following chart illustrates the effect that varying slit sizes have on NIRQuest512 Spectrometer
resolution and throughput:
0
2
4
6
8
10
12
14
16
0
50
100
150
200
Slit Size (microns)
R
e
s
ol
ut
ion (
FW
H
M
)
0
10
20
30
40
50
60
R
el
at
ive I
n
ten
si
ty
NIRQuest Gratings
The following tables show the NIRQuest gratings available for preconfigured (standard) setups and for all
options. Additional grating options, adjustments to starting and ending wavelengths and similar
customization may be available. Please contact an Applications Scientist for details.
Spectrometer
Standard
Grating
Groove
Density
(lines/mm)
Spectral
Range
Blaze
Wavelength
Best Efficiency
(>30%)
NIRQuest512
NIR3
150
~800 nm
1100 nm
900-1700 nm
NIRQuest512-1.9
NIR3
150
~800 nm
1100 nm
1100-1900 nm
NIRQuest512-2.2
NIR2
100
1150 nm
1600 nm
900-2200 nm
NIRQuest512-2.5
NIR1
75
1425 nm
1700 nm
1075-2500 nm
NIRQuest256-2.1
NIR2
100
1150 nm
1600 nm
900-2050 nm
NIRQuest256-2.5
NIR1
75
1425 nm
1700 nm
1075-2500 nm