Ocean Optics Nirquest Install User Manual
Page 35
B: Specifications
016-700000-000-02-201301
29
NIRQuest256-2.1
NIRQuest256-2.5
Defective pixels:
<12 pixels
OPTICAL BENCH
Design:
f/4, symmetrical crossed Czerny-Turner
Entrance aperture
(standard):
25 µm
Entrance aperture (custom
options):
10 µm, 50 µm, 100 µm and 200 µm (or no slit)
Grating options (standard): Grating NIR2, 100 l/mm, 900-
2050 nm
Grating NIR1, 75 l/mm, 1075-2500 nm
Grating options (custom)
2
:
NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13
Longpass filter
3
:
OF1-RG830 longpass NIR filter (optional)
2
nd
Order filter
3
:
Standard
Collimating and focusing
mirrors:
Gold-coated for enhanced NIR reflectivity
Fiber optic connector:
SMA 905 to 0.22 numerical aperture single-strand optical fiber
SPECTROSCOPIC
Wavelength range:
900-2050 nm w/Grating NIR2
900-2500 nm w/Grating NIR1
Optical resolution
(FWHM)
4
:
~7.6 nm w/25 µm slit
~ 9.5 nm w/25 µm slit
Signal-to-noise ratio at full
signal
5
:
10000:1 @ 100 ms integration
7500:1 @ 10 ms integration
A/D resolution:
16-bit
Dark noise:
6 RMS counts @ 100 ms
8 RMS counts @ 10 ms
12 RMS counts @ 250 ms
12 counts RMS @ 30 ms
Dynamic range:
15M (system); 10K:1 for a single
acquisition
500K (system); 7.5K:1 for a single acquisition
Integration time
6
:
1 ms – 2 s
1 ms – 60 ms