beautypg.com

Ocean Optics Nirquest Install User Manual

Page 35

background image

B: Specifications

016-700000-000-02-201301

29

NIRQuest256-2.1

NIRQuest256-2.5

Defective pixels:

<12 pixels

OPTICAL BENCH

Design:

f/4, symmetrical crossed Czerny-Turner

Entrance aperture
(standard):

25 µm

Entrance aperture (custom
options):

10 µm, 50 µm, 100 µm and 200 µm (or no slit)

Grating options (standard): Grating NIR2, 100 l/mm, 900-

2050 nm

Grating NIR1, 75 l/mm, 1075-2500 nm

Grating options (custom)

2

:

NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13

Longpass filter

3

:

OF1-RG830 longpass NIR filter (optional)

2

nd

Order filter

3

:

Standard

Collimating and focusing
mirrors:

Gold-coated for enhanced NIR reflectivity

Fiber optic connector:

SMA 905 to 0.22 numerical aperture single-strand optical fiber

SPECTROSCOPIC

Wavelength range:

900-2050 nm w/Grating NIR2

900-2500 nm w/Grating NIR1

Optical resolution
(FWHM)

4

:

~7.6 nm w/25 µm slit

~ 9.5 nm w/25 µm slit

Signal-to-noise ratio at full
signal

5

:

10000:1 @ 100 ms integration

7500:1 @ 10 ms integration

A/D resolution:

16-bit

Dark noise:

6 RMS counts @ 100 ms

8 RMS counts @ 10 ms

12 RMS counts @ 250 ms

12 counts RMS @ 30 ms

Dynamic range:

15M (system); 10K:1 for a single
acquisition

500K (system); 7.5K:1 for a single acquisition

Integration time

6

:

1 ms – 2 s

1 ms – 60 ms