B: Specifications
016-700000-000-02-201301
31
NIRQuest256-2.1
NIRQuest256-2.5
WIRELESS/ETHERNET INTERFACE (optional add-on accessory)
Wireless (Wi-Fi) interface:
Yes, with adapter
Wired Ethernet interface:
Yes, with adapter
Wi-Fi range:
25 meters in free space
Web server:
Works with common browsers
1
“Useable range” is defined in the context of the NIRQuest model’s detector response and its typical grating response. For example,
the 512-element detector has response at 850 nm, but grating response begins at 900 nm. The G9206 256-element detector
response is sensitive to TEC temperature, and has response only to 2050 nm when the TEC is set to -20 ºC. The G9208 256-
element and 512-element detector has response to 2550 nm, but the grating efficiency drops off at 2500 nm.
2
See
3
Other filter options are available for order-sorting in the NIRQuest256-2.1, NIRQuest256-2.5 and NIRQuest 512-2.5. NIRQuest256-
2.5 and NIRQuest 512-2.5 ship with a 2
nd
-order filter. See an Applications Scientist for details.
4
Optical resolution (FWHM) depends on grating and slit selection.
5
SNR will decrease at longer integration times.
6
Maximum integration times are defined as the longest amount of time one can integrate the spectrometer before the dark level
rises to half of full scale.