1: Introduction

2

891-00000-000-02-201401

The QE Pro’s onboard module has 10 user-programmable digital I/O lines for interfacing to other
equipment; and a pulse generator for triggering other devices. You can use the I/Os to flash a lamp,
stop/start a process, and send a message/alarm during the spectrometer’s integration period. The
spectrometer’s operating parameters can be controlled through software. In fact, wavelength calibration
coefficients unique to each spectrometer are programmed into a memory chip right on the spectrometer.

The QE Pro’s high-speed electronics have been designed for considerable flexibility in connecting to
various modules and external interfaces, including PCs, PLCs and other embedded controllers, through
USB 2.0 or RS-232 communications.

A +5 VDC external power supply is required to operate the spectrometer. This power supply is included
with your spectrometer.

Models

The QE Pro is custom configurable for your specific application. In addition, the following preconfigured
models are offered:

QE Pro-ABS – For absorbance, reflection, irradiance and transmission measurements with a

wavelength range of 200 – 950 nm

QE Pro-FL – For fluorescence measurements with a wavelength range of 350 – 1100 nm
QE Pro-Raman – for 785 nm Raman analysis with a wavelength range of 780 – 1100 nm (preset

for 785 Raman)

Features

• Hamamatsu S7031-1006S Detector:

Typical dynamic range ~85,000:1

Peak QE: 90%

Back-thinned for enhanced UV sensitivity

Integration times from 8 ms to 60 minutes

TE Cooled

• Scientific-grade Optical Bench:

Symmetrical Crossed Czerny Turner

101mm focal length

F number : f/4

Interchangeable slits

14 gratings (H1 – H14); HC1

6 slit widths, plus no slit in SMA or FC bulkhead

• Communications

USB 2.0 Full Speed

RS232 up to 460K Baud

• Thermo Electric Cooler (TEC)

Software-controlled set-point