A: Calibrating the Wavelength of the QE Pro

891-00000-000-02-201401

19

6. Double-click on each of the calibration coefficients displayed in the Wavelength Coefficients

pane of the USB Programmer screen and enter the new values acquired in Step 5 of the

Calibrating the Wavelength of the Spectrometer

section in this appendix.

7. Repeat Step 6 for all of the new values.

8. Click on the Write Coefficients to Device button to save the information, and then select File|

Exit to exit the QE Pro Reflash Utility software.

The new wavelength calibration coefficients are now loaded onto the EEPROM memory chip on the QE
Pro
.