External Triggering Instructions for FW 3.0 and Above

200-00000-001-01-201401

3

Pinout Diagrams for Ocean Optics Spectrometers

HR2000+, Maya2000Pro, QE65000, QE65 Pro, and QE Pro

For external triggering, supply a line from Pin 10 of the multi-pin connector on the spectrometer
to your triggering device. Be sure to also use the Ground Pin when triggering. See the appropriate
spectrometer Data Sheet to locate the ground pin(s).

Pin-out Diagram of 30-pin Connector on HR2000+, Maya2000Pro,

QE65000, QE65 Pro, and QE Pro Spectrometers

USB2000+ and USB4000

For external triggering, supply a line from Pin 7 of the multi-pin connector on the spectrometer to
your triggering device. Be sure to also use Ground Pin 6 when triggering

Pin-out Diagram of 22-pin Connector on USB2000+ and USB4000

Spectrometers

Setting Integration Time in Software

Software, External Hardware Level Trigger and External Hardware Edge Trigger modes can have the
integration time set via OceanView or SpectraSuite.

Set Integration Time with OceanView

Use the Main Controls of the Set Acquisition Parameters in OceanView to select the trigger mode and to
set the integration time. The Automatic button in the Integration Time control box automatically adjusts
the spectrometer’s integration time to bring the highest point of the signal to 85% of full saturation value.
This value is represented by the horizontal blue line on the preview graph.