6 Set up with the display and adjustment module
VEGABAR 82 • Modbus and Levelmaster protocol
46294-EN-131107
Reset - Setup
Menu item
Parameter
Default value
Measurement
loop name
Sensor
Application
Application pro-
cess pressure
Application level
No reset
Slave for elec-
tronic differential
pressure
No reset
Units
Unit of measure-
ment
mbar (with nominal measuring range
≤400 mbar)
bar (with nominal measuring rang-
es ≥1 bar)
Temperature unit °C
Position correc-
tion
0.00 bar
Adjustment
Zero/Min. adjust-
ment
0.00 bar
0.00 %
Span/Max. adjust-
ment
Nominal measuring range in bar
100.00 %
Damping
Integration time
0.0 s
Current output
Current output -
Mode
Output characteristics
4 … 20 mA
Reaction when malfunction occurs
≤ 3.6 mA
Current output -
Min./Max.
3.8 mA
20.5 mA
Lock adjustment
Released
Reset - Display
Menu item
Default value
Menu language
Order-specific
Displayed value 1
Current output in %
Displayed value 2
Measuring cell temperature in °C
Backlight
Switched off
Reset - Diagnosis
Menu item
Parameter
Default value
Sensor status
-
Peak value
Pressure
Actual measured value
Temperature
Actual temperature values from meas-
uring cell, electronics