VEGA VEGABAR 82 Modbus and Levelmaster protocol - Operating Instructions User Manual
Page 32
32
6 Set up with the display and adjustment module
VEGABAR 82 • Modbus and Levelmaster protocol
46294-EN-131107
•
Special characters +, -, /, -
In this menu item you activate/deactivate the slave for electronic dif-
ferential pressure and select the application.
If you have connected a slave sensor, you confirm this with "Activate".
If you have connected no slave sensor, you confirm this with "Deac-
tivate".
VEGABAR 82 can be used for process pressure and level measure-
ment. Default setting is process pressure measurement. The mode
can be changed in this adjustment menu.
The VEGABAR 82 in conjunction with a slave sensor can be used for
flow, differential pressure, density and interface measurement. The
default setting is differential pressure measurement. Switchover is
carried out in the adjustment menu.
Depending on the selected application, different subchapters in the
following adjustment steps are important. There you can find the
individual adjustment steps.
Enter the requested parameters via the appropriate keys, save your
settings with [OK] and jump to the next menu item with the [ESC] and
the [->] key.
In this menu item, the adjustment units of the instrument are deter-
mined. The selection determines the displayed unit in the menu items
"Min. adjustment (zero)" and "Max. adjustment (span)".
Unit of measurement:
If the level should be adjusted in a height unit, the density of the me-
dium must also be entered later during the adjustment.
In addition, the temperature unit of the instrument is specified. The se-
lection determines the displayed unit in the menu items "Peak value,
measuring cell temperature", "Peak value, electronics temperature"
as well as "HART variables".
Setup - Application
Setup - Units