beautypg.com

2 principle of operation – VEGA VEGASON 65 Foundation Fieldbus User Manual

Page 9

background image

l

A

rticle and serial number device

l

A

rticle numbers documentation

l

T

echnical data: For example approvals, process temperature,

process fitting/material, signal output, power supply, protection

W

ith the serial number, you can access the delivery data of the

instrument via www.vega.com, "VEGA Tools" and "serial number
search
". In addition to the type label outside, you can also find the
serial number on the inside of the instrument.

3.2 Principle of operation

VEGASON

65

is an ultrasonic sensor for continuous level measure-

ment. It is particularly suitable for solids, but also good for liquids.

T

he transducer of the ultrasonic sensor transmits short ultrasonic

pulses to the measured product. These pulses are reflected by
product surface and received again by the transducer as echoes. The
running time of the ultrasonic pulses from emission to reception is
proportional to the distance and hence the level. The determined level
is converted into an appropriate output signal and outputted as
measured value.

F

our-wire electronics with separate power supply.

P

ower for the signal output is supplied via the H1 Fieldbus. A two-wire

cable according to Fieldbus specification serves as carrier of both
power and digital data transmission for multiple sensors. This cable
can be operated in two versions:

l

via an H1 interface card in the control system and additional power
supply

l

via a Linking device
with HSE (High speed
E

thernet) and addition-

al power supply ac-
cording to IEC 61158-2

T

he DD (Device Descriptions) and CFF (capability files) necessary for

planning and configuration of your FF (Foundation Fieldbus) commu-
nication network are available in the download area of the VEGA
homepage www.vega.com under "Services - Downloads - Software -

Application area

Functional principle

Power supply and bus
communication

DD/CFF

VEGASON

65 • F

oundation Fieldbus

9

3 P

roduct description

28794

-

EN

-081127