4 mounting, 1 general instructions – VEGA VEGASON 65 Foundation Fieldbus User Manual
Page 12
4 Mounting
4.1 General instructions
M
ake sure that all parts of the instrument in contact with the measured
product, especially the sensor element, process seal and process
fi
tting, are suitable for the existing process conditions such as process
pressure, process temperature as well as the chemical properties of
the medium.
Y
ou can find the specifications in chapter "Technical data" in the or on
the type label.
S
elect an installation position you can easily reach for mounting and
connecting as well as later retrofitting of an indicating and adjustment
module. The housing can be rotated by 330° without the use of any
tools. You can also install the indicating and adjustment module in four
different positions (each displaced by 90°).
U
se the recommended cables (see chapter "Connecting to power
supply") and tighten the cable gland.
Y
ou can give your VEGASON 65 additional protection against
moisture penetration by leading the connection cable downward in
front of the cable entry. Rain and condensation water can thus drain
off. This applies mainly to outdoor mounting as well as installation in
areas where high humidity is expected (e.g. through cleaning
processes) or on cooled or heated vessels.
Fig. 2: Measures against moisture penetration
Suitability for process
conditions
Mounting position
Moisture
12
VEGASON
65 • F
oundation Fieldbus
4 M
ounting
28794
-EN
-081127