beautypg.com

VEGA VEGASON 65 Foundation Fieldbus User Manual

Page 52

background image

45 m (147.64 ft)

30 m (98.425 ft)

15 m (49.213 ft)

90 mm (3.543 in)

6 mm (0.236 in)

-6 mm (-0.236 in)

-90 mm (-3.543 in)

30 mm (1.181 in)

-60 mm (-2.362 in)

60 mm (2.362 in)

-30 mm (--1.181 in)

3 m (9.843 ft)

Fig. 30: Deviation VEGASON 65 versions B, C and D

Influence of the ambient temperature to the sensor electronics

8)

A

verage temperature coefficient of the zero

signal (temperature error)

0.06 %/10 K

Ambient conditions

A

mbient, storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions

P

rocess pressure

-

V

ersion A with PP flange

0

kPa

-

O

ther versions

-20 … 50

kPa/-0.2 … 0.5 bar (-2.9 … 7.3 psi)

P

rocess temperature (transducer temper-

ature)

-40 … +80 °C (-40 … +176 °F)

V

ibration resistance

mechanical vibration with 1 g in the frequency
range 5 … 200 Hz

9)

8)

R

elating to the nominal measuring range.

9)

T

ested according to the regulations of German Lloyd, GL directive 1.

52

VEGASON

65 • F

oundation Fieldbus

10 S

upplement

28794

-EN