VEGA VEGACOM 557 VEGA ASCII protocol User Manual
Page 41
VEGACOM 557 VEGA ASCII
41
High resolution = switch 2.2 = ON
The same enquiry can also be carried out with high resolution. The enquiry to VEGACOM 557
remains the same, however in the answer, more characters are used for the presentation of
the measured value.
Answer of VEGACOM 557
Telegram from processing system to VEGACOM 557
Identifier: 1 character, "%“
T1: 1 character ",“ (comma)
T2: 1 character "L“
End identification: 1 character Chr (13) = CR = Carriage Return
Total length of the telegram: 11 characters
Data image in VEGACOM 557
DCS
DCS
End
S1
Addr
T1
no.
S2
value2
identification
No. of bytes
1
1
1
3
1
7
2
Range (ASCII)
=
0..9
,
001..255
#
DCS value
or fault
%CR
Example
=
2
,
001
#
-000673
%CR
n (n= stated number) repetitions
Start character: S1 = 1 character "=“
Address: Addr = 1 characters VEGACOM 557 address
Separating character: S2 = 1 character ",“ (comma)
Start character: S3 = 1 character "#“
Total length of the telegram: 16 characters
6 digit figure
"-“ with negative values
" " with positive values
valid meas. value
‘FAULT ’
Fault
1 2 3 4 5 6 7
1 2 3 4 5 6 7
DCS
DCS
End
Identifier Address T1
number
T1
number
identification
No. of bytes
1
1
1
3
1
3
1
Range (ASCII)
%
0..9
,
001..255
L
001..255
CR
Example
%
2
,
001
L
2
CR