Veris Technologies VIS_NIR Spectrophotometer - Operating Instructions User Manual
Page 74

Pub# OM17-NIR Shank
5-21
Procedure #7 Replacing halogen lamp
1. Remove the shank from the implement following procedure on page 4-11
2. Remove the top plate on the shank by unscrewing four screws
Figure 22
USE CAUTION WHEN HANDLING TOP COVER! TOO MUCH MOVEMENT
CAN FLEX AND BREAK THE FIBER OPTIC CABLE.
3. Using a 3/32 hex wrench loosen the set screw holding in the lamp housing.
Figure 23