OM17-pH Detector

3-12

Data quality tips
As you collect pH data, monitor the performance of the system by observing the following:
-

During ‘Cycling’ phase or during manual washing, does pH electrode readings get close to the

wash water pH? If not, electrode may not be cleaning properly.
-

During ‘Recording’ phase, does electrode reading move to expected soil pH level?

-Is there any residue in the cavity prepared by the probe chamber?
-Is depth of probe insertion consistent?