7.1 Microphone Calibration 

 

Step 1

Adjust the calibration parameters 

The defaults values are: 

 Averaging time: 

5 s 

 Frequency:   

1 kHz 

 Calibrator Level: 

94 dB 

Step 2

Install the calibrator device on the microphone 

Step 3

Click START 

After the average time is elapsed, the sensitivity value will update. 

Step 4

Click OK to accept the sensitivity value 

 