Fluoview, Fv10i-liv – Olympus FV10i-LIV User Manual
Page 15

12
FLUOVIEW
FV10i-LIV
@
5
Disinfection and Sterilization
}If a hazardous specimen is attached to or splashed in the specimen holder or the incubator, immediately disinfect or
sterilize it.
Do not leave the attached substance, for it will harden and become hard to remove.
Disinfection
· Applicable position: Inside the incubatorand on various specimen holders. (excluding the rubber part @ of the
FV10C-HOW).
· Procedure:
Moisten a clean cloth with disinfecting alcohol and wipe with it.
Do not rub the name plate with a strong force to prevent it from being peeled off.
Sterilization
· Applicable position: Cover on the culture pod.
· Procedure:
Use autoclaving (high-pressure steam sterilization)
at 121°C for 20 minutes.
After sterilization, dry the specimen holder before reuse.