Fluoview, Fv10i-liv – Olympus FV10i-LIV User Manual

Page 15

background image

12

FLUOVIEW

FV10i-LIV

@

5

Disinfection and Sterilization

}If a hazardous specimen is attached to or splashed in the specimen holder or the incubator, immediately disinfect or

sterilize it.
Do not leave the attached substance, for it will harden and become hard to remove.

Disinfection

· Applicable position: Inside the incubatorand on various specimen holders. (excluding the rubber part @ of the

FV10C-HOW).

· Procedure:

Moisten a clean cloth with disinfecting alcohol and wipe with it.
Do not rub the name plate with a strong force to prevent it from being peeled off.

Sterilization

· Applicable position: Cover on the culture pod.
· Procedure:

Use autoclaving (high-pressure steam sterilization)
at 121°C for 20 minutes.
After sterilization, dry the specimen holder before reuse.