7

Fig. 2

Connecting the culture perfusion tubes to the culture pod

Connecting the chemical stimulation tube to the culture pod

Be sure to connect to the OUT side make the amount of the culture solution 2 ml in order to prevent
bubble production in the culture solution.

Culture pod

Tube

Inner dia. 2 x Outer dia. 4 mm

Chemical syringe

CAUTION

4

²

³

1. Remove the cap @ provided on the left side of main body (Fig. 3).

Use the allen wrench included in the product. Loosen the stopper screw
from inside and remove the cap by bringing the stopper in horizontal
position.

2. Remove the cover ² of the incubator.
3. Remove the screw ³ by the allen wrench and remove the cap | provided

on the side of the cover (Fig. 2).

4. Connect the tubes to culture pod as illustrated in the Fig. 3.
}Connect the tube on the side of the peristaltic pump to OUT.

Connect the tube on the side of culture solution to IN.

5. Place the cover on the incubator. Use the space for the cap | to take out

the tubes.

Culture pod

Peristaltic
pump

Cap @

Tube
Inner dia. 2 x Outer
dia. 4 mm

Culture solution

OUT

IN

FV10C-W3

Incubator

Fig. 3