Fluoview, Fv10i-liv – Olympus FV10i-LIV User Manual
Page 13
10
FLUOVIEW
FV10i-LIV
Fig. 9
· Loosen the knob | of the slide glass stopper and lift it.
· Place the well slide 5 along the left edge, lower the slide glass stopper
and tighten the knob |.
6
7
Remove media chamber from the slide and cover glass on top
surface. Shield it with proper method.
Place the slide with cover glass facing downward for imaging.
}When using the specimen holder for slide glass, memorize the map area
indicator number 7 aligned with the position of the observation target 6
and use the same position in later observation/acquisition (Fig. 9).
4. Hold the specimen holder ² horizontally by the holder handle 8 or
pillars of the specimen stoppers and place the specimen holder on the
specimen holder mount so that the positioning hole 9 of the specimen
holder comes on the front (Fig. 7).
5. Re-place the cover of the incubator.
6. Push down the close lever a and confirm that a click sound is generated
(Fig. 7).
}If the cover is not closed completely, the internal temperature of the
incubator may become unstable and the laser beam will not be output,
making observation impossible.
Fig. 8
4
5
Well slide specimen (Fig. 8)