Fluoview, Fv10i-liv – Olympus FV10i-LIV User Manual

Page 13

background image

10

FLUOVIEW

FV10i-LIV

Fig. 9

· Loosen the knob | of the slide glass stopper and lift it.
· Place the well slide 5 along the left edge, lower the slide glass stopper

and tighten the knob |.

6

7

Remove media chamber from the slide and cover glass on top
surface. Shield it with proper method.
Place the slide with cover glass facing downward for imaging.

}When using the specimen holder for slide glass, memorize the map area

indicator number 7 aligned with the position of the observation target 6
and use the same position in later observation/acquisition (Fig. 9).

4. Hold the specimen holder ² horizontally by the holder handle 8 or

pillars of the specimen stoppers and place the specimen holder on the
specimen holder mount so that the positioning hole 9 of the specimen
holder comes on the front (Fig. 7).

5. Re-place the cover of the incubator.
6. Push down the close lever a and confirm that a click sound is generated

(Fig. 7).

}If the cover is not closed completely, the internal temperature of the

incubator may become unstable and the laser beam will not be output,
making observation impossible.

Fig. 8

4

5

Well slide specimen (Fig. 8)