Draining of Water Pump (Fig. 6)

The water pump for the objective lens contains water to prevent dryness. At the first startup of FV10i-LIV, or after emptying
and refilling the objective water tank with water, drain water from the water pump according to the procedures below.

1. Open the body cover and remove the sample holder.

2. Start Software and Logon.

3. Click on the

button. Water feeding starts and the color

of the button changes to red.

4. Wait for the color of the

button to return automatically to

gray.

5 minutes later, water feeding stops and the color of the button returns

automatically to gray.

5. Return the sample holder and the body cover back to the original position.

Fig. 6

4

Placing the Specimen

(Figs. 7 to 9)

· If you attempt to place the specimen directly on the specimen holder installed in the main unit, it may drop

inside the main unit.
Be sure to take out the specimen holder and place the specimen on it outside the main unit.

· If the sample falls within the device, pick up the sample with the tweezers made of bamboo or plastic, or the

stick wrapped by the adhesive tape.
If you open the cover, the temperature within the incubator drops. When you replace samples, etc., do so
quickly. After you open the cover, wait for a while to stabilize the temperature within the incubator.
(When the use environment temperature is 25°C and the cover open time is less than 3 minutes, wait for
approx.30 minutes.)

Place a dummy specimen until the internal temperature of the incubator rises to the optimum temperature (35 to
37°C).
When the specimen holder for 35 mm dish is used, place dummy specimens in all of the three positions.
Without specimens placed, the internal temperature of the incubator may become unstable.

Fig. 7

1. Push and hold the cover buttons @, and lift the cover.
2. Remove the cover of the incubator.
3. Place the specimen.
} When using a culture pod, it is necessary to remove its top cover, set the

35 mm dish and connect the tubes (see page 7).

Specimen other than well slide/culture pod specimen

· Place the specimen in the specified specimen holder ² and fix the

specimen with the specimen stopper ³ (Fig. 7).
A slide glass specimen should be placed with the cover glass facing
downward.

²

³

9

CAUTION

NOTE